LABORATORY CASE STUDY • BIOMAGNETISM

Ultra-Low Noise Dynamic Field Compensation for Wearable Brain Imaging (MEG/MCG)

Wearable magnetoencephalography (MEG) and magnetocardiography (MCG) using Optically Pumped Magnetometers (OPMs) represent a revolution in non-invasive clinical neuroscience and cardiology.

Unlike conventional cryogenic SQUID systems that require rigid, liquid-helium dewars, microfabricated OPM vapor cells can be placed directly on the patient’s scalp or chest. However, operating unshielded or lightly-shielded OPM sensors in realistic clinical environments requires dynamic magnetic field cancellation of extraordinary purity.

The Metrology Challenge: Sensor Sensitivity vs. Driver Noise

State-of-the-art rubidium and cesium OPMs achieve field sensitivities below 10 to 15 fT/√Hz. When large electromagnetic coils (Helmholtz, Lee-Whiting, or bi-planar coil arrays) are placed around a seated or moving patient to null Earth’s ambient ~50 μT field and 50/60 Hz power-grid interference, the current driving those coils must not introduce electronic noise that overwhelms the minute biomagnetic signals of the human brain or fetal heart.

The Metrology Threshold: A current source driving a 10 nT/mA compensation coil must exhibit current noise well below 10 pA/√Hz (and ideally ∼2 pA/√Hz) across the 0.1–100 Hz biomagnetic band to ensure the synthetic field noise floor remains beneath 10 fT/√Hz.

The DM Technologies Solution

Leading biomagnetism groups deploy the Multichannel Current Source (MCS) Low-Range Version specifically configured for OPM active shielding:

Performance Metric Standard Range Low-Range OPM Configuration
Full Scale Output ±4 mA / ±40 mA per channel ±0.4 mA / ±4 mA per channel
Current Resolution (LSB) 0.125 μA / 1.25 μA 12.5 nA / 125 nA (16-bit monotonic)
Current Noise Density 10 pA/√Hz @ 1 kHz ∼2 pA/√Hz @ 10 kHz (< 5 pA/√Hz in MEG band)
Output Compliance Voltage ±10 V ±10 V (inductive coil stable with 100 nF filter)

Clinical & Neuroscience Implementations

Institutions such as PTB Berlin, the University of Nottingham, and ICFO Barcelona utilize DM Technologies multi-channel current sources to drive multi-axis magnetic compensation systems for:

  • Wearable Epilepsy Focal Mapping: Recording ictal and inter-ictal discharges in pediatric patients who cannot remain motionless in traditional cryogenic MEG scanners.
  • Fetal Magnetocardiography (fMCG): Non-invasive monitoring of fetal cardiac arrhythmia from 24 weeks gestation, completely unhindered by maternal vernix caseosa.
  • Cognitive Motor Neuroscience: Measuring neural motor responses during natural head and body movement inside actively-compensated light shields.